The EZTest Biological Test Pack is a small, convenient, FDA
510(k) cleared
product containing EZTest self-contained biological indicators. Designed to parallel the performance of the AAMI Test Pack, each pack includes a record card with a chemical integrator.
EZTest Biological Test Packs (PCDs) are available in a pouch for EO sterilization, and a box (about the size of a deck of playing cards) for steam sterilization.
Availability
| Steam Biological Indicator Test Pack (PCD) - 25 units per case |
| Spore |
Incubation |
Dimensions |
Population |
Reorder # |
| G. stearothermophilus |
24 Hours at 55-60°C |
4-1/2" x 2-5/8" x 3/4" |
105 |
STP/05 (5 controls)
STP/25 (25 controls) |
| Ethylene Oxide Biological Indicator Test Pack (PCD) - 25 units per case |
| Spore |
Incubation |
Dimensions |
Population |
Reorder # |
| B. atrophaeus |
48 Hours at
35-39°C |
Pouch: 3-1/2" x 9-1/4" |
106 |
ETP/05 (5 controls)
ETP/25 (25 controls) |
Use (steam)
- Load sterilizer.
- Lay EZTest biological test pack flat on lowest shelf.
- Process load with a normal cycle.
- Remove the test pack – allow to cool at room temperature for 15 minutes.
- Remove the EZTest biological indicator and incubate at 55-60°C for 24 hours.
Use (EtO gas)
- Load sterilizer.
- Lay EZTest biological test pack flat in the center of the sterilizer load.
- Process load with a normal cycle.
- Remove the test pack as soon as possible from the load.
- Remove the EZTest biological indicator from the test pack and incubate at 35-39°C for 48 hours.
Storage
Room temperature.
Expiration
24 months from date of manufacture.
Certification
EZTest biological indicators are traceable to a recognized culture collection and certified for population, D-value, Z-value and performance.